6482-901-01 Rev. A / August 2012
Return to
16-27
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 16: SCPI Command Reference
Read display
DATA?
:DISPlay[:WINDow[1]]:DATA?
Read top display
:DISPlay:WINDow2:DATA?
Read bottom display
Description
These query commands are used to read what is currently being displayed on the
top and bottom displays. After sending one of these commands and addressing
the Model 6482 to talk, the displayed data (message or reading) will be sent to the
computer.
Define :TEXT messages
DATA <a>
:DISPlay[:WINDow[1]]:TEXT:DATA <a>Define message; top display
:DISPlay:WINDow2:TEXT:DATA <a>Define message; bottom display
Parameters
<a> = ASCII characters for message
Types:
String
‘aa...a’ or “aa...a”
Indefinite Block #0aa...a
Definite Block
#XYaa...a
where
Y = number of characters in message:
Up to 20 for top display
Up to 32 for bottom display
X = number of digits that make up Y (1 or 2)
Query
:DATA?
Query the defined text message
Description
These commands define text messages for the display. A message can be as
long as 20 characters for the top display, and up to 32 characters for the bottom
display. A space is counted as a character. Excess message characters result in
an error.
An indefinite block message must be the only command in the program message
or the last command in the program message. If you include a command after an
indefinite block message (on the same line), it will be treated as part of the mes-
sage and is displayed instead of executed.
STATe <b>
:DISPlay[:WINDow[1]]:TEXT:STATe <b>Control message; top display
:DISPlay:WINDow2:TEXT:STATe <b>Control message; bottom display
Parameters
<b> =
0 or OFF
Disable message for specified display
1 or ON
Enable message for specified display
Query
:STATe?
Query state of message mode for specified display
Description
These commands enable and disable the text message modes. When enabled, a
defined message is displayed. When disabled, the message is removed from the
display.
GPIB operation
: A user defined text message remains displayed only if the
instrument is in remote. Taking the instrument out of remote (by pressing the
LOCAL key or sending LOCAL 25) cancels the message and disables the text
message mode.
RS-232 operation
: A user defined test message can be cancelled by sending the
:SYSTem:LOCal command or pressing the LOCAL key.