6482-901-01 Rev. A / August 2012
Return to
11-5
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 11: Output Enable and Output Configuration
NOTE
Output enable and source clear commands are global and affect both channels
simultaneously.
Output configuration programming example
lists the command sequence for output configuration. These commands set up the
Model 6482 as follows:
•
Output enable: On
•
Auto-off mode: On
NOTE
Connect pins 8 and 9 of the output enable connector to each other to simulate a closed
output enable switch. If you do not connect pins 8 and 9, the instrument will not turn on its
output when the measurement is made.
Table 11-1
Output configuration commands
Command
Description
:OUTPut[1]:ENABle <state>
:OUTPut[1]:ENABle:TRIPped?
:SOURce[1]:CLEar
:SOURce[1]:CLEar:AUTO <state>
:SOURce[1]:CLEar:AUTO:MODE <name>
:SOURce2:CLEar
:SOURce2:CLEar:AUTO <state>
:SOURce2:CLEar:AUTO:MODE <name>
Enable or disable output enable (state = ON or OFF).
Query output enable tripped state (1 = Tripped).
Turn output source off when in idle state.
Enable or disable automatic output off. State = ON (output off
after measurement) or OFF (output stays on).
Automatic clear mode. Name = ALWays (every reading;
default) or TCOunt (ON when trigger layer entered; OFF
when leaving trigger layer).
Turn output source off when in idle state.
Enable or disable automatic output off. State = ON (output off
after measurement) or OFF (output stays on).
Automatic clear mode. Name = ALWays (every reading;
default) or TCOunt (ON when trigger layer entered; OFF
when leaving trigger layer).
Table 11-2
Output configuration programming example
Command
Description
*RST
:FORM:ELEM CURR1
:SOUR1:VOLT 10
:OUTP:ENAB ON
:SOUR1:CLE:AUTO ON
:READ?
Restore GPIB defaults.
Select channel 1 reading data.
Output 10 V on channel 1.
Activate output enable.*
Enable automatic off (auto-off) mode.
Trigger and acquire readings.
*Connect pins 8 and 9 of the output enable connector to simulate closed output
enable switch.