6482-901-01 Rev. A / August 2012
Return to
17-21
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 17: Calibration
Step 4. Channel 2 voltage source calibration adjustment
Follow the steps below to calibrate both channel 2 bias voltage ranges.
summarizes
calibration ranges and voltages.
1. Connect the DMM to the Model 6482 channel 2 INPUT and OUTPUT connectors, as shown in
. Connect the DMM INPUT HI to the VOLTAGE SOURCE OUTPUT CHANNEL 2
connector; connect the DMM INPUT LO to the inner shield of the INPUT CHANNEL 2
connector.
2. Select the DMM DC volts function, and enable autorange.
3. From normal display, press the Model 6482
SRC2
key, and then set the channel 2 voltage bias
source to the 10 V range using the RANGE
key.
4. Press the
MENU
key, select
CAL
and press the
ENTER
key.
5. Select
EXECUTE
and press the
ENTER
key to enter the CAL EXECUTION menu.
6. Select
V-CAL
and press the
ENTER
key. The instrument will display the following:
V-CAL CHANNEL SELECT
CHAN-1 CHAN-2
7. Select
CHAN-2
and press the
ENTER
key. The instrument will prompt for +10 V calibration:
V-CAL
2
Press ENTER to 10.000 V
8. Press the
ENTER
key. The instrument will display the following message:
DMM RDG: 10.0000 V
Use
,
,
,
, ENTER or EXIT.
9. Note the DMM voltage reading, and then use the EDIT keys to adjust the Model 6482 display
value to match the noted reading.
10. Press the
ENTER
key. The instrument will prompt for the 0 V calibration point:
V-CAL
2
Press ENTER to Output 00.000 V
11. Press the
ENTER
key. The instrument will display the following message:
DMM RDG: 00.00000 V
Use
,
,
,
, ENTER or EXIT.
12. Note the DMM voltage reading, and then use the EDIT keys to adjust the Model 6482 display
value to match the noted reading.
13. Press the
ENTER
key. The instrument will display the prompt for the -10 V calibration point:
V-CAL
2
Press ENTER to Output -10.000 V
14. Press the
ENTER
key. The instrument will display the following message:
DMM RDG: -10.00000 V
Use
,
,
,
, ENTER or EXIT.
15. Note the DMM voltage reading, and then use the EDIT keys to adjust the Model 6482 display
value to match the noted reading and press the
ENTER
key.
16. Press the
EXIT
key to return to normal display.
17. Using the appropriate RANGE key, select the 30 V bias voltage range.
18. Repeat steps 4 through 15 for the 30 V bias voltage range.
Содержание 6482
Страница 3: ......
Страница 62: ...Section 4 Range Digits Speed and Filters In this section Topic Page Range and digits 4 2 Speed 4 4 Filters 4 6...
Страница 90: ...Section 7 Sweep Operation In this section Topic Page Sweep types 7 2 Configuring and running a sweep 7 5...
Страница 182: ...Section 14 Common Commands In this section Topic Page Command summary 14 2 Command reference 14 2...