17-30
Return to
6482-901-01 Rev. A / August 2012
Section 17: Calibration
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Step 7. Enter calibration adjustment dates and save calibration
NOTE
For temporary calibration without saving new calibration constants, proceed to
.
1. From the CALIBRATION menu, select
SAVE
, and then press the
ENTER
key. The instrument
will prompt you for the calibration date:
CAL DATE: 06/01/2012
Use
,
,
,
, ENTER or EXIT.
2. Using the EDIT keys, change the displayed date to today’s date and press the
ENTER
key.
Press the
ENTER
key again to confirm the date.
3. The instrument will then prompt for the calibration due date:
NEXT CAL: 06/01/2012
Use
,
,
,
, ENTER or EXIT.
4. Set the calibration due date and press the
ENTER
key. Press the
ENTER
key again to confirm
the date.
5. When the calibration dates are entered, calibration is complete and the following message will
be displayed:
CALIBRATION COMPLETE
Press ENTER to save; EXIT to abort
6. Press the
ENTER
key to save the calibration data or press the
EXIT
key to abort without saving
calibration data.
Step 8. Lock out calibration
From the CAL EXECUTION menu, select
LOCK
and press the
ENTER
key to lock out calibration.
Press the
EXIT
key to return to normal display.
Remote calibration
Use the following procedure to perform remote calibration by sending SCPI commands over the
IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters are separately
summarized for each step.
Remote calibration command summary
summarizes only those remote calibration commands used in this section.