6482-901-01 Rev. A / August 2012
Return to
7-7
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 7: Sweep Operation
The source delay is part of the source-delay-measurement (SDM) cycle and is used to allow the
source to settle before the measurement is made. See
for details.
The total time period of the source delay could include an autodelay and a user programmed
delay. With autodelay enabled, 1 ms of delay is used. The user-programmable source delay adds
0000.0000 to 9999.9980 seconds of delay. See
,
to set these delays.
Additional delay for a sweep is available by using the trigger delay. This user-specified delay
(0000.0000 to 9999.9990 seconds) occurs before each SDM cycle (device action) of the sweep.
The trigger delay is executed before each new source-point in the sweep. See
panel operation of the trigger model”
to set trigger delay.
Trigger count and sweep points
The trigger count and number of sweep points should be either the same or multiples of one
another. For example, with five sweep points and a trigger count of 10, the sweep will run twice.
See
for details on trigger count.
Performing sweeps
Procedures for the various sweep types are covered below.
NOTE
The following procedure assumes that the Model 6482 is already connected to the device
under test (DUT) as explained in
.
Performing a linear staircase sweep
Step 1. Configure the channel.
Configure the Model 6482 for a specific channel as follows:
1. Select the source function by pressing the
SRC1
or
SRC2
key.
2. Set the appropriate source level value.
3. Press the
MSR1
or
MSR2
key to select the measurement channel, and then select the
measurement range.
for more information.
The source level you set becomes the bias level for the sweep. When turned on, the output will
maintain this bias level until the sweep is started. Typically, 0 V is used as the bias level.
If using a fixed measurement range, either make sure it can accommodate every measurement
point in the sweep or use autoranging.
Step 2. Configure the sweep.
Configure the sweep as follows:
1. Press the
SRC1
or
SRC2
key to select the channel, and then press the
EXIT
key to leave edit
mode.
2. Press the
CONFIG
key and then the
SWEEP
key.
3. Select
TYPE
, and then press the
ENTER
key.
4. Select
STAIR
, and then press the
ENTER
key.
5. At the prompts, enter the START, STOP, and STEP values.
6. From the CONFIGURE SWEEPS menu, select
SWEEP-COUNT
, and then press the
ENTER
key.