How do I optimize performance?
There are three primary factors that affect measurement accuracy and speed:
•
Warm-up:
For rated measurement accuracy, allow the Model 2651A to warm up for at least two
hours before use.
•
Speed setting:
The Speed setting affects both speed and accuracy (see
(on page
•
Autozero:
Autozero can be disabled to increase speed at the expense of accuracy (see
Disabling autozero to increase speed
Setting speed
To set speed from the front panel:
Press the
SPEED
key, and then choose the setting based on your requirements:
•
Choose the
FAST
setting for the fastest measurements (note that measurement accuracy will be
reduced).
•
For best accuracy, use the
HI-ACCURACY
setting (note that speed will be reduced).
•
To balance accuracy and speed, use the
MED
or
NORMAL
setting.
For additional information, including the OTHER setting and remote configuration, refer to
Disabling autozero to increase speed
In certain measuring situations, disabling autozero (setting it to OFF) can increase measurement
speed. If autozero is disabled, accuracy will drift with time and temperature.