Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-336
2651A-901-01 Rev. A / March 2011
Details
The
trigger.blender[
N
].reset()
function resets the following attributes to their factory defaults:
•
trigger.blender[
N
].orenable
•
trigger.blender[
N
].stimulus[
M
]
It also clears
trigger.blender[
N
].overrun
.
Also see
(on page 7-334)
trigger.blender[N].stimulus[M]
(on page 7-336)
trigger.blender[N].stimulus[M]
This attribute specifies which events trigger the blender.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Trigger blender N reset
Not saved
0
Usage
eventID
= trigger.blender[
N
].stimulus[
M
]
trigger.blender[
N
].stimulus[
M
] =
eventID
eventID
The event that triggers the blender action; see
Details
N
An integer representing the trigger event blender (
1
to
4
)
M
An integer representing the stimulus index (
1
to
4
)
Details
There are four acceptors that can each select a different event. The
eventID
parameter can be the event ID of
any trigger event.
Use zero to disable the blender input.
The
eventID
parameter may be one of the existing trigger event IDs shown in the following table.
Trigger event IDs*
Event ID
Event description
smu
X
.trigger.SWEEPING_EVENT_ID
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
smu
X
.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
smu
X
.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smu
X
.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measure action