Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-194
2651A-901-01 Rev. A / March 2011
Also see
(on page 7-177)
(on page 7-179)
(on page 7-181)
(on page 7-211)
(on page B-16)
smuX.measure.count
This attribute sets the number of measurements performed when a measurement is requested.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall
Saved setups
1
Usage
count
= smu
X
.measure.count
smu
X
.measure.count =
count
count
Number of measurements.
X
SMU channel (for example,
smua.measure.count
applies to SMU Channel A).
Details
This attribute controls the number of measurements taken any time a measurement is requested. When using a
reading buffer with a measure command, this attribute also controls the number of readings to be stored.
If
smu
X
.measure.count
is set to a value greater than 1, any measurement delay set by
smu
X
.measure.delay
will only occur before the first measurement, while the
smuX.measure.interval
controls the interval between successive measurements.
Example
smua.measure.count = 10
Sets the SMU Channel A measure count to 10.
Also see
(on page 7-201)
(on page 7-209)