Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-368
2651A-901-01 Rev. A / March 2011
Also see
tspnet.timeout
This attribute sets the timeout value for the
tspnet.connect()
,
tspnet.execute()
, and
tspnet.read()
commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset Not saved
20.0
Usage
value
= tspnet.timeout
tspnet.timeout =
value
value
The timeout duration in seconds (0.001 s to 30.000 s)
Details
This attribute sets the amount of time the
tspnet.connect()
,
tspnet.execute()
, and
tspnet.read()
commands will wait for a response.
The time is specified in seconds. The timeout may be specified to millisecond resolution, but is only accurate to
the nearest 10 ms.
Example
tspnet.timeout = 2.0
Sets the timeout duration to two seconds.
Also see
(on page 7-365)
tspnet.tsp.abort()
This function stops remote instrument execution.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
tspnet.tsp.abort(
connectionID
)
connectionID
Integer value used as a handle for other tspnet commands
Details
Sends an abort command to the remote instrument.