Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-199
smuX.measure.highcrangedelayfactor
This attribute contains a delay multiplier that is used during range changes when High-C mode is active.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall
Saved setups
10
Usage
delayFactor
= smu
X
.measure.highcrangedelayfactor
smu
X
.measure.highcrangedelayfactor =
delayFactor
delayFactor
The delay factor. Set to a value between 1 and 99.
X
SMU channel (for example,
smua.measure.highcrangedelayfactor
applies
to SMU Channel A).
Details
This delay multiplier is only active when High-C mode is active.
Example
smua.measure.highcrangedelayfactor = 5
Increases the delay for SMU channel A by a factor if 5.
Also see
(on page 7-209)
(on page 7-215)
smuX.measure.interval
This attribute sets the interval between multiple measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall
Saved setups
0
Usage
interval
= smu
X
.measure.interval
smu
X
.measure.interval =
interval
interval
The interval value (in seconds). Set to a value between 0 and 1.
X
SMU channel (for example,
smua.measure.interval
applies to SMU
Channel A).