Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-232
2651A-901-01 Rev. A / March 2011
Usage
eventID
= smu
X
.trigger.endpulse.stimulus
smu
X
.trigger.endpulse.stimulus =
eventID
eventID
Set to the event that triggers the end pulse source off action
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.stimulus
applies to SMU channel A)
Details
Set this attribute to the event ID of any trigger event generator to wait for that event. To bypass waiting for an
event, set this attribute's value to
0
. Set
eventID
to one of the existing trigger event IDs shown in the following
table.
Trigger event IDs*
Event ID
Event description
smu
X
.trigger.SWEEPING_EVENT_ID
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
smu
X
.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
smu
X
.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smu
X
.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measure action
smu
X
.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smu
X
.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smu
X
.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a digital I/O line
tsplink.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
lan.trigger[
N
].EVENT_ID
Occurs when the appropriate LXI trigger packet is
received on LAN trigger object
N
display.trigger.EVENT_ID
Occurs when the TRIG key on the front panel is
pressed.
trigger.EVENT_ID
Occurs when a *TRG command is received on the
remote interface.
(GPIB only) Occurs when a GET bus command is
received.
(VXI-11 only) Occurs with the VXI-11 command
device_trigger
. Reference the VXI-11 standard
for additional details on the device trigger operation.
trigger.blender[
N
].EVENT_ID
Occurs after a collection of events is detected.
trigger.timer[
N
].EVENT_ID
Occurs when a delay expires.
* Use the name of the trigger event ID to set the stimulus value rather than the numeric value. Using the name
makes the code compatible for future upgrades (for example, if the numeric values must change when
enhancements are added to the instrument).
Example
smua.trigger.endpulse.action =
smua.SOURCE_IDLE
smua.trigger.endpulse.stimulus =
trigger.timer[1].EVENT_ID
Configure end pulse action to achieve a
pulse and selects the event
trigger.timer[1].EVENT_ID
that will
cause the arm event detector to enter the
detected state.