![Keithley 2651A Скачать руководство пользователя страница 372](http://html1.mh-extra.com/html/keithley/2651a/2651a_reference-manual_661729372.webp)
Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-64
2651A-901-01 Rev. A / March 2011
display.locallockout
This attribute describes whether or not the EXIT (LOCAL) key on the instrument front panel is enabled.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Power cycle
Not saved
display.UNLOCK
Usage
lockout
= display.locallockout
display.locallockout =
lockout
lockout
0
or
display.UNLOCK
: Unlocks EXIT (LOCAL) key
1
or
display.LOCK
: Locks out EXIT (LOCAL)
key
Details
Set
display.locallockout
to
display.LOCK
to prevent the user from interrupting remote operation by
pressing the
EXIT (LOCAL)
key.
Set this attribute to
display.UNLOCK
to allow the EXIT (LOCAL) key to interrupt script/remote operation.
Example
display.locallockout = display.LOCK
Disables the front-panel EXIT (LOCAL) key.
Also see
None
display.menu()
This function presents a menu on the front panel display.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
selection
= display.menu(
name
,
items
)
selection
Name of the variable that holds the menu item selected
name
Menu name to display on the top line
items
Menu items to display on the bottom line