Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-329
Details
Data for current measurements, voltage source values, and timestamps are stored in
smuX.nvbuffer1.
If all parameters are omitted when this function is called, this function is executed with the parameters set to the
default values.
Performs a linear voltage sweep with current measured at every step (point):
1. Sets
the
smu
to output
startv
amperes, allows the source to settle for
stime
seconds, and then
performs a current measurement.
2. Sets
the
smu
to output the next amperes step, allows the source to settle for
stime
seconds, and
then performs a voltage measurement.
3. Repeats the above sequence until the voltage is measured on the
stopv
amperes step.
The linear step size is automatically calculated as follows:
step = (stopv – startv) / (points – 1)
Figure 108: SweepVLinMeasureI()
startv
= Current measurement
stopv
stime
stime
stime
Example
SweepVLinMeasureI(smua, -1, 1, 1E-3, 1000)
This function performs a 1000-point linear
voltage sweep starting at -1 V and stopping at
+1 V. Current is measured at every step (point)
in the sweep after a 1 ms source settling period.
Also see
None
SweepVListMeasureI()
(on page 5-22) function performs a voltage list sweep with current measured at every
step (point).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function No