Section 3: Functions and features
Model 2651A High Power System SourceMeter® Instrument Reference Manual
3-68
2651A-901-01 Rev. A / March 2011
Enabling high-capacitance mode
Before enabling high-capacitance mode, note the following:
•
It is important to read the previous section to understand the impact of high-capacitance mode.
•
Test the DUT and the capacitor to determine the best current source and range of output voltages.
•
The settling times can vary based on the DUT. It is important to test the limits of the DUT before you
use high-capacitance mode.
•
Failure to test the DUT for the appropriate current source and output voltages can result in damage to or
destruction of the DUT.
•
For optimal performance, do not continuously switch between normal mode and high-capacitance
mode.
•
Before you charge the capacitor, start with 0 (zero) voltage across the capacitor.
•
When high-capacitance mode is enabled, a minimum load capacitance of 100 nF is recommended. In
absence of this minimum load capacitance, overshoot and/or ringing may occur.
Front panel
To enable high-capacitance mode from the front panel:
1. Press
the
CONFIG
key, then select
SRC
>
HIGHC-MODE
.
2. Select
SRC-ENABLE
>
ENABLE
. High-capacitance mode is enabled.
3. Push
the
ENTER
key.
4. Press
the
EXIT (LOCAL)
key to back out of the menu structure.
Command interface
Turning on High-C mode has the following effects on the SMU settings:
•
smua.measure.autorangei
is set to
smua.AUTORANGE_FOLLOW_LIMIT
and cannot be
changed.
•
Current ranges below 1
μ
A are not accessible.
•
If
smua.source.limiti
is less than 1
μ
A, it is raised to 1
μ
A.
•
If
smua.source.rangei
is less than 1
μ
A, it is raised to 1
μ
A.
•
If
smua.source.lowrangei
is less than 1
μ
A, it is raised to 1
μ
A.
•
If
smua.measure.lowrangei
is less than 1
μ
A, it is raised to 1
μ
A.