Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-148
2651A-901-01 Rev. A / March 2011
Figure 104: PulseIMeasureV
level
bias
ton
bias
toff
toff
Voltage measurement
Example
PulseIMeasureV(smua, 0.001, 1.0,
20E-3, 40E-3, 10)
SMU A will output 1 mA and dwell for 40 ms, output
1 A and dwell for 20 ms. The voltage measurement
occurs during the 20 ms dwell periods. After the
measurement, the output will return to 1 mA and
dwell for 40 ms. This pulse-measure process will
repeat nine more times.
Also see
None
PulseVMeasureI()
(on page 5-23) function performs a specified number of pulse V, measure I cycles.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function No
Usage
PulseVMeasureI(
smu, bias, level, ton, toff, points
)
smu
System SourceMeter
®
instrument channel (set to
smua
)
bias
Bias level in volts
level
Pulse level in volts
ton
Pulse on time in seconds
toff
Pulse off time in seconds
points
Number of pulse-measure cycles