Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2651A-901-01 Rev. A / March 2011
3-79
Bit
B8
B7
B6
B5
B4
B3
B2
B1
Annunciator
SMPL
STAR
TRIG
ARM
AUTO
4W
MATH
FILT
Weighted
value*
128
64 32 16 8 4 2 1
Binary
value 0/1 0/1 0/1 0/1 0/1 0/1 0/1 0/1
* The weighted values are for bits that are set to “1.” Bits set to “0” have no value.
Not all of the above indicators shown in above table may used by the Model 2651A.
Local lockout
The front-panel EXIT (LOCAL) key is used to cancel remote operation and return control to the front
panel. However, this key can be locked-out to prevent a test from being interrupted. When locked, this
key becomes a NO-OP (no operation). Configure the following attribute to lock or unlock the
EXIT (LOCAL) key:
display.locallockout =
lockout
Where
lockout
is set to one of the following values:
0
or
display.UNLOCK
1
or
display.LOCK
Example:
The following programming example illustrates how to lock out the EXIT (LOCAL) key:
display.locallockout = display.LOCK
Load test menu
The LOAD TEST menu lists tests (USER, FACTORY, and SCRIPTS) that can be run from the front
panel. Factory tests are preloaded and saved in nonvolatile memory at the factory. They are available
in the FACTORY TESTS submenu. Named scripts that have been loaded into the run-time
environment can be selected from the SCRIPTS submenu. Refer to
(on page 6-3) for
additional information.
User tests
User tests can be added or deleted from the to the USER TESTS submenu.