12 - 8
ADM 6725 E10
KAT
Wall Thickness Measurement
•
Disconnect the plug of the probe cable
from the FLUXUS socket.
12.5.8 Conventions when in Super-User Mode
The general rules for the Super-User Mode are also applicable for the WTM
option.
This means:
•
No limit values for all input parameters
•
Display of measured values that are outside the specified limits
•
Probe parameters can be edited in program branch PARAMETER
12.6 Program Branch WTM-OUTPUT OPTIONS
FLUXUS modifies the program branch OUTPUT OPTIONS for the operational
mode wall thickness measurement (WTM).
(However, this cannot be recognised explicitly.)
12.6.1 Set-up of the Measuring Quantity
Output Options
par mea>OPT<sf
The selection of the quantity of measurement
can be defined, as with flow measurement, in
program branch OUTPUT OPTIONS.
Meas.
Quantity
Wall thickness
The wall thickness of a material will be measured
by using the known longitudinal sound velocity,
which must be entered in program branch
PARAMETER.
Meas.
Quantity
c-LONGITUDINAL
The longitudinal sound velocity will be measured
by using the known thickness of the material,
which must be entered in program branch
PARAMETER.
Attention!
The wall thickness measurement always determines the longitudinal sound velocity.
However, for the flow measurement, the transversal sound velocity of the pipe
material* is usually used. This means:
With the WTM option, the sound velocity cannot always be determined for
'Other Material' required for flow measurement.
* With some exceptions, the transversal sound velocity for a material is
about 30 % ... 60 % of the longitudinal sound velocity.
12.6.2 Set-up of other Output Options
It is possible to print or transmit online the measured values of the WTM option
(wall thickness or sound velocity). This might be useful for, e.g. documenting all
conditions of a measuring data set.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Страница 4: ...ADM 6725 E10 KAT ...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally ...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop ...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally ...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally ...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...