ADM 6725 E10
KA
T
4 - 7
Program Branch PARAMETER
Attention!
The roughness may be entered within the range from 0 to 5 mm. Otherwise, an error
message appears (see page 4-1).
4.3 Input of Medium Parameters
Medium parameters required for measurement are:
•
Minimum and maximum sound
velocity of the medium
•
Density of the medium
(only if output option
mass flow is activated)
•
Kinematic viscosity of the medium
•
Temperature of the medium
(Table 4 in chapter 17 gives an overview of pre-programmed parameters for those media which are often used.)
Medium
"
P11 Select from this scroll display
group the medium you want to
measure.
Confirm by pressing
ENTER
.
#
Water P11-1
#
Petrol P11-2
#
Ether P11-3
#
Methanol P11-4
#
Acetone P11-5
#
Toluene P11-6
#
HCL 37%
P11-7
#
Glycol P11-8
#
Vinylchloride P11-9
#
Milk P11-10
#
BP Transcal LT
P11-11
#
BP Transcal N
P11-12
#
R22 Freon
P11-13
#
CaCl2 -45 C
P11-14
#
CaCl2 -15 C
P11-15
#
R134 Freon
P11-16
#
H20-Ethan.-Glyc. P11-17
#
Oil P11-18
#
Ammonia (NH3)
P11-19
#
Shell Thermina
P11-20
!
Other Medium
P11-21
Next display
P16
(Without pressing ENTER,
the displays scroll around)
If the medium has been selected, FLUXUS jumps straight to the display for
entering the
medium temperature
. If the list does not contain your lining material,
the set-up can be done as follows:
Other Medium
Medium
"
P11-21 Scroll through to the display
'Other
Medium'
.
Confirm by pressing
ENTER
.
Now, FLUXUS requests the input for the medium parameters min.
and max. sound velocity, kinematic viscosity and density.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Страница 4: ...ADM 6725 E10 KAT ...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally ...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop ...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally ...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally ...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...