ADM 6725 E10
KA
T
12 - 5
Wall Thickness Measurement
Attention!
Temperature fluctuations change the sound velocity of the material. The zero
adjustment of the probe using a comparative block should be performed at the
measuring location to minimise the effect of temperature fluctuations.
12.5.2 Measurement of the Sound Velocity
c-LONGITUDINAL
m/s?
M* This display appears after the calibration, if
sound velocity was selected as the quantity
of measurement. When there is no valid
measuring value, the unit of measurement
and a question mark are shown on the
lower display line.
c-LONGITUDINAL
√
5370 m/s
M* Firmly press the probe bearing a film of the
acoustic coupling component against the
test piece (the pipe) for the measurement of
its sound velocity. As soon as a valid
measuring value has been obtained, a tick
sign and the sound velocity are shown.
12.5.3 Further Information when Measuring
O N
O
O F F
O
With these keys, the displays showing
further information about the measurement
status can be switched ON or OFF.
The meaning of the messages is as follows:
NO COUPLING
Wall thickness
M* No signal can be detected.
•
Either there is no coupling component,
•
or the input value for the sound
velocity in program branch
PARAMETER is inaccurate.
MATER.
TOO THICK
Wall Thickness
M* The thickness of the test piece (the pipe) is
greater than the upper measuring limit of
the probe.
MATER. TOO THIN
Wall Thickness
M* The thickness of the test piece (the pipe) is
thinner than the lower measuring limit of the
probe.
ERROR SIGNAL
Wall thickness
F* FLUXUS cannot process the signal of the
probe.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Страница 4: ...ADM 6725 E10 KAT ...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally ...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop ...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally ...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally ...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...