14 - 1
Errors
14 Errors
This chapter contains an overview of error messages you might encounter when operating
FLUXUS. We will describe their causes, give possible reasons for their occurrence and
also try to make suggestions as to how these problems can be overcome.
14.1 Error Messages
14.1.1 Errors during Parameter Input
Error messages F1 and F2:
25.2 MINIMUM
Outer Diameter
Reason:
The entry for the outside diameter was smaller than
25.2 mm or greater than 3100 mm.
Corrective action:
3100.0 MAXIMUM
Outer Diameter
Make sure your inputs are within the respective limit
values. Should the pipe parameters be outside the
given limits, measurements are not possible under
normal circumstances.
Please consult your distributor for advice.
Error message F3:
Too Small !
Inner Diameter
Reason:
The inner diameter, which is calculated from the
entries for wall thickness and outside diameter, is
smaller than the minimum value allowed.
(It could be
possible that you have corrected the outside pipe diameter without taking
the wall thickness into consideration.)
Corrective action:
Check the entered parameters of outside pipe
diameter and wall thickness.
Note:
The error messages F1 to F3 do not appear in Super-User Mode.
Error message F4:
NO DATA !
par >MEA< opt sf
Reason:
There are parameters missing in program branch
PARAMETER, no complete parameter record exists.
It could be that you have not worked through the
program branch PARAMETER.
Corrective action:
Enter any parameters not yet entered.
Error message F5:
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Страница 4: ...ADM 6725 E10 KAT ...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally ...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop ...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally ...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally ...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...