Technical Data
ADM 6725 E10
KA
T
16 - 1
16 Technical Data
Specification subject to alterations without notice
Measuring instrument
Housing: Aluminium
powder-coated
Front panel:
Polycarbon material
(see chapter 17, table 5)
Size:
H 118 x W 276 x D 310 mm
(with handle)
Weight: 3.5
kg
Display:
2 x 16 characters, dot matrix, backlit
Power supply:
Internal rechargeable NiCd-battery, 6V/4 Ah, or external power supply, 14
h operating time with internal battery
Power consumption:
< 2.5 W in measuring mode
Ambient temperature:
-10 ° ... 60 °C
Degree of protection:
IP 54 according to EN 60529
Clamp-on flow transducers
Transducer type M2N, M2E
Transducer type Q3N, Q3E
Rated
(possible)
diameter range
DN (50) 100 ... DN 6500
DN (10) 25 ... DN 400 (1000)
Dimensions
60 x 30 x 34 mm
43 x 18 x 22 mm
Material
Stainless steel
Stainless steel
Temperature range
M2N:
-30 °C ... 130 °C
M2E:
-30 °C ... 200 °C, for short
periods up to 300°C
Q3N:
-30 °C ... 130 °C
Q3E:
-30 °C ... 200 °C, for short
periods up to 300°C
Degree of protection
IP 65 acc. to EN 60529, IP68
optional
IP 65 acc. to EN 60529, IP68
optional
Transducer cable
In flexible stainless steel conduit;
1 m, 2 m, 5 m + DN-range
(extension up to 50 m possible)
In flexible stainless steel conduit;
1 m, 2 m, 5 m + DN-range
(extension up to 50 m possible)
Mains power adapter and battery charging unit
Housing: Plastic
material
Size:
100 x 70 x 200 mm
Power supply:
230 V AC / 50 Hz
Output voltage:
9 V DC
Output current:
Max 1.2 A
Units of measurement:
Volume flow
Flow velocity
Mass flow
Total quantity
Heatflow
Fluid velocity
Volume
Mass
Heatflo
w
Heat
quantity
m3/h m/s
g/s m3 g
W
J,
Wh
m/s
m3/min inch/s
t/h
l
kg
kW
kJ,
kWh
m3/s
kg/h
gal t
MW
MJ,
MWh
l/h
kg/min
l/min
l/s
ugph
bls/d
1 gallon [US] = 3.78 l; 1 barrel = 42 gallons = 158.76 l
Transfer of measured values
Memory capacity:
•
27,000 values, optional > 100,000 values
•
Max. 100 measuring data sets
•
Max. 14 different parameter records; 8o memory locations
Communication: RS232
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Страница 4: ...ADM 6725 E10 KAT ...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally ...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop ...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally ...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally ...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...