10 - 4
ADM 6725 E10
KA
T
Virtual Calculation Channels
10.5 Measuring with Calculation Channels
par>MEA<opt sf
Measuring
After selecting program branch >MEASURING< ...
Menu in case of a 2-
channel unit:
CH: A B>Y<Z
ON:
√
√
√
-
Menu in case of a 3-
channel unit:
CH: A B C>Y<Z
ON:
√
√
.
√
.
WARNING! CHANNEL
B:INACTIVE
... It is possible to activate the calculation channels in the
same way as the physical channels. The menu displayed
depends on the number of physical channels.
The ticks denote which of the different channels are active.
Should no physical channel be selected the unit automatically
returns to the main menu.
Should the required physical channel be selected as an input
into the virtual channel but not activated, this warning
message appears.
...
Now, the positioning of the transducers for all activated
physical channels begins.
Y:Volume Flow
-53.41 m3/h
Y: A - B
-53.41 m3/h
In order to display the data produced through the virtual
channels the unit has to operate in MANUAL-MUX mode (key
1). This is installed automatically once a calculation channel
has been activated, and the information is displayed.
Whilst measurement is taking place it is possible to view the
calculation being carried-out by pressing the DISP-Key.
10.6 Process Outputs from the Calculation
Channels
The allocation of process outputs to measuring channels (Special Functions\System
Settings\ Process Outputs) is the same for both physical and virtual channels.
SYSTEM settings
Proc. outputs
Example:
If you want to output the totalisation (quantity measurement) of
a calculation channel via a binary output by pulses.
Select SYSTEM settings in program branch SPECIAL
FUNCTIONS. Install the binary output B1 as pulse output.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Страница 4: ...ADM 6725 E10 KAT ...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally ...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop ...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally ...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally ...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...