Technical Data
16 - 2
ADM 6725 E10
KA
T
Measuring characteristics
Rated inside diameter:
Depending on transducer type
Flow velocity range:
0.01 m/s ... 25 m/s
Flow volume:
(See enclose flow volume nomogram at the end of this chapter)
Resolution: 0.025
cm/s
Repeatability:
0.15 % of measured value
±
0.015 m/s
Accuracy:
Volume flow:
±
1 ... 3 % of measured value depending on application.
±
0.5 % of measured value with process calibration.
Flow velocity:
±
0.5 % of measured value
The stated accuracy figures can only be achieved, if pipe and medium parameters are known and entered
correctly, and an axi-symetrical flow profile is fully developed, i.e. that the recommendations for the minimum
distances between measuring point and flow disturbance sources have been observed.
Recommended distance to
flow disturbance sources:
Straight length upstream
10 x D ... 50 x D
Straight length downstream
5 x D ... 10 x D
D = rated pipe diameter at measuring point (chapter
3.4)
Gaseous or solid content: < 10 % of volume
Cut-off flow
Adjustable, acc. to a flow velocity of 0 ... 12.7 cm/s
Signal damping:
0 ... 60 s, user configurable
Response time
1 s, optional 70 ms
Measuring cycle
100 ... 1000 Hz, single channel
Calculation functions
Average/difference/sum
Process interfaces (optional)
Process outputs
Galvanically isolated from main electronics
- Current 0/4 ... 20 mA
Passive:
(U
ext
< 24 V
Active:
(R
ext
< 500
Ω
- Voltage
0 ... 1 V or 0 ... 10 V, R
i
= 500
Ω
- Frequency
0 ... 1 kHz, or 0 ... 10 kHz; (OC)
- Digital (pulse, status)
Totalizer valued: 0.01 ... 1000 /unit; width: 80 ... 1000 ms; (OC/Reed)
Reed=Reed-NO contact (300 V/ 0.5 A); OC=Open-Collector
- Relays (alarm outputs)
max. 3 channels; individually programmable; (Reed/OC/TP)
Ree
d
= Reed relay contact
(48 V/0.5 A)
OC
= Open-Collector
(36 V/4 mA)
TP
= Totem-Pool
(5 V)
Process inputs
Galvanically isolated from main electronics
- Temperature:
PT 100, four-wire circuit; measuring range: -50 °C ... 400 °C
- Current:
0 ... 20 mA; R
i
= 50
Ω
- Voltage:
0 ... 1 V; R
i
= 1 M
Ω
;
Wall thickness measurement
Measuring range
1.0 mm ... 200 mm
Resolution 0.01
mm
Linearity 0.1
mm
Wall thickness probe DA301, HT400
Measuring range
1.0 ... 200 mm
Resolution 0.01
mm
Linearity 0.1
mm
Temperature range
Standard version DA301: -20 °C ...+60 °C
High temperature version HT400: 0 ...+200 °C, for short
periods up to +540 °C
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Страница 4: ...ADM 6725 E10 KAT ...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally ...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop ...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally ...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally ...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...