Process Interfaces
ADM 6725 E10
KA
T
13 - 9
A relay of
TYPE: NON HOLDING
will switch for about
1 second
, if the limit is exceeded
for a short time or the flow direction sign changes.
Relay contacts, which have been programmed
FUNC: QUANTITY
, will
immediately
switch as soon as the current total flow reaches or exceeds the limit.
Relay contacts, which have been program-
med with
FUNC: CHANGE-SIGN
(+
$
$
$
$
-
$
$
$
$
-+)
, will switch for about
1 second
with any change of flow direction,
(Relay
type: NON HOLDING
).
However, a relay
type: HOLDING
will
switch with the first change of flow
direction. It may be switched back by this
key.
+
-
Relay output Type: HOLDING
Relay output Type: NON HOLDING
0
ca. 1 s
manual switching
of the relay
Flow by change-sign
Relay contacts, which have been programmed
FUNC: ERROR
, will only switch after
several unsuccessful measuring attempts (SIGNAL-lamp lights red). Therefore, typical
short-term disturbances of the measurement in the form of air bubbles, e.g. caused by
pumps being switched on, will not trigger the relay.
If there is an internal adaptation to changing measuring conditions, e.g. to a considerable
rise of the medium temperature, the relay will not switch.
As soon as a measured value is obtained (SIGNAL-lamp lights green), a relay TYPE:
NON HOLDING will switch back again.
If you have selected
Function: NONE
, the relay will automatically be reset to switch
setting: NO Contact. This means that the relay coil is de-energised (contacts are open).
Note:
FLUXUS doesn't indicate via visual or acoustic means whether the relays switch or
reset.
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725 ...
Страница 4: ...ADM 6725 E10 KAT ...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally ...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally ...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop ...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally ...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally ...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally ...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally ...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally ...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally ...