Guidance and manufacturer’s declaration – electromagnetic immunity
9619NB/9626NB(-2)/9627NB(-2)
is intended for use in the electromagnetic environment speci
Ƃ
ed below. The customer or user of the
9619NB/9626NB(-2)/9627NB(-2)
should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic Discharge (ESD)
IEC 61000-4-2:
|
2008
± 6 kV contact
± 8 kV air
Complies
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4:
|
2004
± 2 kV for power supply lines
± 1 kV for input/output lines
Complies
± 2 kV for power supply lines
± 1 kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5 :
|
2005
± 1 kV differential mode
± 2 kV common mode
Complies
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycle
<40% U
T
(60% dip in U
T
)
for 5 cycles
<70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Complies
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycle
Complies
<40% U
T
(>60% dip in U
T
)
for 5 cycles
Complies
<70% U
T
(>30% dip in U
T
)
for 25 cycles
Complies
<5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the equipment or system requires
continued operation during power mains interruptions,
it is recommended that the equipment or system be
powered from an uninterruptible power supply.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8:
|
2004
3 A/m
Complies
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.