Example of Climate Chamber
Installation
0507-016
Figure 90
Climate Chamber Installation
(not Available from JUMO)
The following numbers refer to Figure 90:
1
=
PTFE sleeve
2
=
Cable gland, for example: AGRO 1100.12.91.065
3
=
Stainless steel cable tie or similar fastener
l
4
=
To be sealed (silicone)
5
=
Temperature probe
6
=
Relative humidity probe
7
=
HMP247CG, Cable gland AGRO (available from JUMO)
NOTE
Let the cables hang loosely to prevent condensed water running to the
probe.