Chapter 7 _____________________________________________________________ Technical Data
JUMO _________________________________________________________________________ 151
Dewpoint Temperature (907023/337
Warmed Probe Option)
Find the intersection of the dewpoint temperature curve and the
dewpoint difference reading (process temperature-dewpoint
temperature) on the x-axis and read the accuracy in dewpoint
measurement on the y-axis.
0508-017
Figure 72
Accuracy in Dewpoint Measurement
Operating Environment
Operating temperature range
for humidity measurement
-70 ... +180 °C (-94 ... +356 °F) see probe
specifications
for transmitter body electronics
-40 ... +60 °C (40 ... +140 °F)
with display
0 ... +60 °C (+32 ... +140 °F)
Storage temperature
-55 … +80 °C (-67 … +176 °F)
Electromagnetic compatibility
EN61326-1:1997+ Am1:1998
+
Am2:2001 Industrial environment