Chapter 5 _______________________________________________________________ Maintenance
JUMO _________________________________________________________________________ 133
Error States
In error state the quantity is not measured and the output is shown as
follows:
- Analog channel outputs 0 mA or 0 V (you can use the serial line
command
AERR
or display/keypad to change this fault indication
value, see section Analog Output Fault Indication Setting on page
112.)
- The serial port outputs stars (***)
- The cover LED is blinking
- Optional display: error indicator is lit.
0706-008
Figure 67
Error Indicator and Error Message
The following number refers to Figure 67 oben:
1
= Error Indicator
-
The error indicator disappears when the error state is over and you
have checked the error message. Press the
INFO
button to display
the error message.
You can also check the error message via the serial interface by using
the command
ERRS
. In case of constant error, please contact JUMO.