Chapter 4 _________________________________________________________________ Operation
JUMO __________________________________________________________________________ 89
Where the readings (from the left) are:
24.1720 = Temperature of the humidity probe ( ºC)
15.0399 = RH (%RH)
-3.5743 = Tdf (C)
189.2324 = Capacitance (pF)
15.0709 = RH raw: calculated from scaled capacitance (%RH)
15.0399 = Enhancement factor corrected RH (%RH)
23.9765 = Temperature of the additional temperature probe
(optional) (ºC)
Formatting Serial Line Message
FTIME and FDATE
FTIME
and
FDATE
commands will enable/disable output of time
and date to the serial line. To add time to R and SEND outputs enter:
FTIME
[
x
]
To add date to
R
and
SEND
outputs enter:
FDATE
[
x
]
where
x
=
ON or OFF
Example:
>send
RH= 98.4 %RH T= 31.0 'C
>ftime on
Form. time : ON
>send
03:47:59 RH= 98.4 %RH T= 31.0 'C
>fdate on
Form. date : ON
>send
2004-07-05 03:48:03 RH= 98.4 %RH T= 31.0 'C
>
FST
To output the state of optional probe heating and chemical purge in
connection with SEND and R commands enter:
FST
[
x
]