or for additional T probe:
CTA
4.
Type
C
and press
ENTER
a few times to check if the reading is
stabilized. Let the reading stabilize, give the reference
temperature after the question mark and press
ENTER
three
times
.
When using two reference temperatures (2-point calibration)
press
ENTER
only twice and insert the probe to the second
reference. When the reading is stabilized, give the second
reference temperature after the question mark and press
ENTER
. Please, note that the difference between the two
temperature references must be at least 30 ºC.
Example (1-point adjustment):
>ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>
5.
OK
indicates that the calibration has succeeded. Enter the
calibration information (date and text) to the transmitter's
memory; see the serial commands
CTEXT
and
CDATE
.
6.
Press the
ADJ
button on the motherboard to close the
adjustment mode.
7.
Take the probe out of the reference conditions and replace the
filter.