User's Guide _______________________________________________________________________
122 ______________________________________________________________________________
INTV
Use the
INTV
command to set the RUN mode output interval.
INTV
[
n xxx
]
where
n
=
1 - 255
xxx
=
S, MIN or H
This command sets the RUN mode output interval. The time interval
is used only when the RUN mode is active. For example, the output
interval is set to 10 minutes.
>INTV 10 min
Output intrv. :
10 min
>
Setting RUN output interval to zero enables the fastest possible output
rate.
ADDR
Addresses are required only for POLL mode (see serial line command
SMODE on page 97). Use the
ADDR
command to input the RS-485
transmitter address.
OPEN [
aa
]
where
aa
=
address (0 ... 99) (default = 0)
Example:
the transmitter is configured to address 99.
>ADDR
Address : 2 ? 99
>