OS1 Mobile Operating Chair
Operating / Service Manual
8
COMMISSIONING
3 Commissioning
The chapter commissioning describes the preparation of the chair for use including a final check of the functional
safety.
The use of specially trained and instructed service personnel during commissioning is neither necessary
or intended
3.1 Safety Instructions for Commissioning
Before commissioning the chair, the user must be instructed in the handling of the chair using the instruction man-
ual, having previously read it in detail. In addition, the potential dangers that may occur despite proper operation
of the chair must be pointed out in detail.
Operator’s obligation - instruction
• The operator must observe the respective valid national specifications and legal regulations
• The operator must ensure that the users are instructed before using the chair
• The instruction can be given either by an authorized person of an authorized specialist distributor or
directly by the manufacturer
TIP
If the delegated person or the operator, e.g. the person responsible for medical devices, has been
instructed by an authorized person of an authorized specialist distributor or by the manufacturer,
we would like to point out that we authorize the person responsible for medical devices of the
operation within the scope of the purpose for the instruction of the chair on the basis of training,
experience and knowledge.
• The operator must ensure that the instruction manual has been read and understood completely
by the users. The operator must disclose the place where the instruction manual is stored and make
it accessible at all times.
• The operator must ensure that representative personnel also receive adequate instruction in the
operation of the chair
• The operator must ensure that the chair is operated exclusively by trained users
• The operator must inform the users that the patients may only operate the chair after their
instruction, if necesary only under supervision
TIP
Electrical testing before commissioning
A further electrical inspection of the chair before the first commissioning is not necessary, as the
chair has been tested for electrical safety in accordance with EN 60601 (test report in instruction
manual) and has left our factory in perfect condition. In other countries, the applicable national
regulations and laws must be observed. Additional commissioning specifications of the operator
must always be observed.
Danger of suffocation from packaging material
Packaging material is not a toy and must be kept away from babies and small children.
In particular, do not pull the plastic bags or sacks over your head or crawl into them.
DANGER