11
OS1 Mobile Operating Chair
Operating / Service Manual
COMMISSIONING CONTINUED
3.5 Requirements on the Place of Operation
For the operation of the chair, an appropriate mains supply and, if applicable, an equipotential bonding connection
is required at the place of use. If this is available and the building’s wiring permits it, the chair must always be
connected to the equipotential bonding system.
Battery
TIP
Charge the battery for at least 24 hours before first use. The battery is then available for
normal operation without restriction.
3.6 Check Before Commissioning and Reassignment
After the commissioning work has been completed, and before any occupation by a new patient, the chair must
be cleaned and disinfected (See Cleaning and Disinfection).
CAUTION
Risk of infection from contaminated chair
The chair must be cleaned and then disinfected before the first use. This also applies before
each use (occupation by a new patient).
WARNING
Risk of injury for patient, user, and third parties
The chair must be checked for functional safety after first commissioning and every occupation
by a new patient. It must be ensured that the chair can be used as intended without endangering
the patient, the user, or third parties.
If the chair is put into operation for the first time or before any occupation by a new patient, the same functional
check must be carried out as for recommissioning after storage.
• General check of the overall condition of the chair for soiling, condition and damage, completeness and
legibility of stickers, symbols, and instructions, clean if necessary in accordance with the manufacturer’s
specifications or those of the operator and repair if necessary.
• Visually inspect the mains cable and plug, hand control, as well as all other electrical cables and
connectors for breakage, proper installation (e.g. strain relief), clamping or friction points etc. and
repair if necessary.
• Functional check of the electrical adjustment devices and the control system. All electric actuators must
be operated once up to their end positons. Pay particular attention to changed operation behavior,
unusual noises, speed, smooth running, odors and increased temperature and repair if necessary.
• Carry out a functional check according to the installed brake system for the casters (locking, freewheel)
and repair if necessary.
• Any defects or damage found must be remedied immediately. The chair must not be used before the
defects or damage have been remedied.
WARNING
Risk of injury due to defective chair
The chair must
not
be operated in a defective or faulty condition in which the chair could endanger
patients, users, or third parties. A defective or faulty chair must be clearly marked
“DEFECT”
.
The marking must be done in such a way that the defective condition is clearly identifiable
for everyone. Inform the operator for intensive inspection during repair.