Chapter 6 DTF Testing
CellAdvisor JD720C Series User
’s Manual
55
Figure 25
Frequency/distance setup window for DTF measurements
Procedure
1
Press the
SETUP
hard key and touch the
Frequency/Distance
tab.
2
Do one of the following:
–
To set the start and stop frequencies, complete the following steps:
a
Select the cell next to the
Start Frequency
.
b
Enter a value by using the numeric keys or the rotary knob.
c
Select the cell next to the
Stop Frequency
.
d
Enter a value by using the numeric keys or the rotary knob.
–
To set the center frequency and the span, complete the following steps:
a
Select the cell next to the
Center Frequency
.
b
Enter a value by using the numeric keys or the rotary knob.
c
Select the cell next to the
Span
.
d
Enter a value by using the numeric keys or the rotary knob.
–
To select a band from the band list, complete the following steps:
a
Touch the icon
next to the band list.
The Band List window appears. It displays a list of standard bands stored in the
instrument and custom bands created in the JDViewer. See
b
Highlight your choice from the list, and then tap the
Apply
button on the Band List
window.
The instrument automatically updates the values in the
Start Frequency
and
Stop
Frequency
cells with the selected band's start and stop frequencies.
3
Tap the
Apply
button in the lower-right corner of the setup screen to save the change(s).
NOTE
Frequency changes after calibration affect the calibration status. If changed within the
registered band, the status turns into CAL ON (I) and it does not require re-calibration. If
changed off the registered band, the status turns into CAL OFF and it requires
re-calibration.