5
Right-click your CellAdvisor product under Devices and select
Connect using
and then
Direct connection
.
The message
“Connection successful” appears on your computer. You can also check the
Bluetooth connection on your JD720C series by going to
System > Bluetooth
.
6
See
“Setting up communication in JDSU application” on page 181 for more information on how to
set up JDViewer, JDRemote, or other JDSU application.
Setting up communication in JDSU application
If you have successfully connected your JD720C series to your computer to use the
Bluetooth Personal Area Network
service, you can configure communication setting on the
JDSU application.
JDViewer setup
Procedure
1
Launch the JDViewer application.
2
Go to
Settings > Communication
.
The Settings dialog box appears.
3
Select the
Communication
tab, click the drop-down list in the Instrument pane, and then select
the
JD720C Series
from the list.
4
In the Connection Method pane, select the
TCP/IP
, and then enter
192
.
168
.
10
.
10
.
5
Click
OK
to finish the communication setup.
6
Go to
Load > Connect
to make a connection between the instrument and JDViewer.
JD720C series and JDViewer become ready for communication.