Chapter 18 System Setting and Maintaining
CellAdvisor JD720C Series User
’s Manual
159
NOTE
Older version of firmware does not have the
Server
option. It is
recommended that you update your firmware with the
USB
option first.
b
Enter
ftp2.jdsu.com
in the server address field.
c
Enter your proxy server address in the proxy server field, if necessary.
Contact your system administrator for your proxy server information.
d
Tap the
Check
button to check.
e
If the installed firmware version is the latest one, the message
“
Your firmware is up to
date.
” is displayed. If not, the instrument displays the latest firmware version information
and you can install it by touching the
Install
button.
NOTE
If the instrument encounters power down during the upgrade, it may not operate properly.
Do not interrupt power supply until the upgrade is done.
Performing service diagnostic
The Service Diagnostic enables you to view system log, run self-testing, revert to factory
settings, and perform touch-screen alignment.
Procedure
1
Press the
MODE
hard key and then tap the
System
button.
The System window appears.
2
Touch the
Service Diagnostic
tab.
The Service Diagnostic window appears as shown in
3
Use the following action buttons as needed:
–
–
–
–
See the sections below for details of each task.
4
Tap the
Apply
button in the lower-right corner of the screen.