Preparing for fiber inspection
Connecting a fiber microscope
Connecting an optional fiber microscope P5000i is mandatory for fiber inspection and
analysis. See
“Fiber microscope kits” on page 170 for available options.
Procedure
1
Connect your fiber microscope P5000i to one of the
USB Host
ports of the JD720C series.
The instrument detects the connected video microscope automatically.
The
Fiber Inspection
icon and name appears on the measurement mode screen.
Selecting measurement mode
Procedure
1
Press the
MODE
hard key and then select the
Single Layout
tab.
2
Double-tap the
Fiber Inspection
icon.
3
Optional
. To check the connected fiber microscope such as model number, version, and serial
number, complete the following steps:
a
Press the
SETUP
hard key and then touch the
Info/Setting
tab.
The information and setting window appears.
b
Tap the
Cancel
button to exit.
Setting parameters
You can set testing parameters such as analysis profile, tip, test button, and auto center. You
can also select the profile and the tip from the drop-down lists activated on the
Live
screen.
Procedure
1
Press the
SETUP
hard key and touch the
Info/Setting
tab.
The information and setting window appears.
2
To select one of the pre-configured analysis profiles that match the PASS/FAIL criteria in the IEC
visual inspection standard, complete the following steps:
a
Tap the drop-down list button for the
Profile
.
b
Select the profile that you want to use, from the following choices:
–
MM (IEC-61300-3-35)
–
Ribbon, MM (IEC-61300-3-35)
–
Ribbon, SM APC (IEC-61300-3-35)
–
SM APC (IEC-61300-3-35)
–
SM PC (IEC-61300-3-35)
–
SM UPC (IEC-61300-3-35)
NOTE
All of the profiles are factory set and cannot be edited or removed.