Chapter 4 Performing Calibration
CellAdvisor JD720C Series User
’s Manual
41
O-S-L-Thru calibration using electronic EZ-Cal kit
Procedure
1
Connect an extension cable, to the
RF Out / Reflection
port of the instrument.
2
Connect the
EZ-Cal
to the end of the extension cable connected to the instrument's the
RF Out /
Reflection
port.
3
Connect the USB cable to the side of the
EZ-Cal
and then to the
USB Host
port to power the
e-calibration kit.
4
Press the
CAL
hard key.
The on-screen instruction appears to guide you through the e-calibration.
5
Tap the
O-S-L-Thru
button in the lower-left corner of the screen.
The calibration progress bar appears.
6
Disconnect the
EZ-Cal
kit and then connect the end of the extension cable connected to the
instrument
’s
RF Out / Reflection
port to the instrument
’s
RF In
port as illustrated in
7
Tap the
Continue
button in the lower-left corner of the screen.
The calibration progress bar appears.
8
Check the screen to see if the calibration status is now changed to
CAL ON (E)
.