![JDS Uniphase CellAdvisor JD720C Series Скачать руководство пользователя страница 142](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd720c-series/celladvisor-jd720c-series_user-manual_3625723142.webp)
Chapter 17 Operation
142
CellAdvisor JD720C User's Manual
display with two enabled zoom zones.
6
Tap the
Apply
button to save the changes and exit.
Figure 92
Quick save setup window
Using save on event
The
Save On Even
enables you to automatically save an active trace or screen according to
your configuration set in Save On Event in advance if a measurement result falls outside
defined limit setting.
Procedure
1
Press the
SAVE/LOAD
hard key.
The Save/Load screen appears.
2
Touch the
Save On Event
tab.
The Save On Event window appears as shown in
3
In the file manager pane on the left-hand side, highlight a folder or location in the internal memory
or in your USB memory drive, to which you want to save.
4
To name your file, complete the following steps:
a
Touch the
File Name
box.
The on-screen keyboard appears.
b
Type a file name in the
Name
field by touching keys on the on-screen keyboard.
See "Using on-screen keyboard" on page 146 for details.
c
Tap the
Save
button in the lower-right corner of the keyboard screen.
The on-screen keyboard dismisses.
5
Select
On
for the
Save On Event
.
6
Select the
Save Type
option from the choices:
Screen
(
*.png
) and
Trace
(
*.tra
).
7
Select the
Save Then Stop
option from the following choices:
–
On:
Saves one failure event and then hold the measurement. If you see the text
“Hold” on
the screen, pressing the
HOLD