![JDS Uniphase CellAdvisor JD720C Series Скачать руководство пользователя страница 157](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd720c-series/celladvisor-jd720c-series_user-manual_3625723157.webp)
Chapter 18 System Setting and Maintaining
CellAdvisor JD720C Series User
’s Manual
157
NOTE
The USB Test & Measurement Class (USBTMC) is a standard for programmatic control of
USB-based test instruments that defines protocols used to send and receive messages.
Figure 106
Instrument settings window
Installing options
You can install your purchased options, e.g., Bluetooth Connectivity, USB GPS Connectivity.
Procedure
1
Press the
MODE
hard key and then tap the
System
button.
The System window appears.
2
Touch the
Instrument Settings
tab.
The Instrument Settings window appears as shown in
3
Tap the
Installable Options
button for the
License
.
The Installable Options window appears.
4
Enter a license number you have received at your purchase of the option.
5
Tap the
Apply
button to save.
Configuring network
You can configure the Ethernet network to connect your JD720C series and your PC with
JDViewer or JDRemote installed.
Procedure
1
Press the
MODE
hard key and then tap the
System
button.
The System window appears.
2
Touch the
Instrument Settings
tab.