Pro I: Analog Input Modules
Pro-AIn-32/12 Rev. B
ADwin
50
ADwin-Pro
Hardware, manual version 2.9, June 2006
Setting the Input Voltage Range
The input module
is equipped with an ADC whose input
voltage range is adjustable by 2 DIL switches. As a default setting, the ADC is
set to the voltage range of ±10V. The settings for other voltage ranges can be
found in
Fig. 78 –
: Board and front panel
After every jumper setting you have to recalibrate the ADC, in order to assure
correct measurement results. The adjustment of gain and offset is made by
software. The individual steps are described in the
.
Fig. 76 –
assignment single-ended
Fig. 77 –
: Pin
assignment differential
ANALOG IN 17
ANALOG IN 18
ANALOG IN 19
ANALOG IN 20
ANALOG IN 21
ANALOG IN 22
ANALOG IN 23
ANALOG IN 24
ANALOG IN 25
ANALOG IN 26
ANALOG IN 27
ANALOG IN 28
ANALOG IN 29
ANALOG IN 30
ANALOG IN 31
ANALOG IN 32
AGND
EVENT IN
ANALOG IN 1
ANALOG IN 2
ANALOG IN 3
ANALOG IN 4
ANALOG IN 5
ANALOG IN 6
ANALOG IN 7
ANALOG IN 8
ANALOG IN 9
ANALOG IN 10
ANALOG IN 11
ANALOG IN 12
ANALOG IN 13
ANALOG IN 14
ANALOG IN 15
ANALOG IN 16
AGND
RESERVED
DGND
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
37
36
35
34
33
32
31
30
29
28
27
26
25
24
23
22
21
20
ANALOG IN 1 (-)
ANALOG IN 2 (-)
ANALOG IN 3 (-)
ANALOG IN 4 (-)
ANALOG IN 5 (-)
ANALOG IN 6 (-)
ANALOG IN 7 (-)
ANALOG IN 8 (-)
ANALOG IN 17 (-)
ANALOG IN 18 (-)
ANALOG IN 19 (-)
ANALOG IN 20 (-)
ANALOG IN 21 (-)
ANALOG IN 22 (-)
ANALOG IN 23 (-)
ANALOG IN 24 (-)
AGND
EVENT IN
ANALOG IN 1 (+)
ANALOG IN 2 (+)
ANALOG IN 3 (+)
ANALOG IN 4 (+)
ANALOG IN 5 (+)
ANALOG IN 6 (+)
ANALOG IN 7 (+)
ANALOG IN 8 (+)
ANALOG IN 17 (+)
ANALOG IN 18 (+)
ANALOG IN 19 (+)
ANALOG IN 20 (+)
ANALOG IN 21 (+)
ANALOG IN 22 (+)
ANALOG IN 23 (+)
ANALOG IN 24 (+)
AGND
RESERVED
DGND
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
37
36
35
34
33
32
31
30
29
28
27
26
25
24
23
22
21
20
19AD0102
A0 A1 A2 A3 A4 A5 A6 A7
74LS
19
ON
DIP
1
2
3
4
5
6
7
8
IN
A
105
MA
X
6225
D
S
1867
C1
-1
00
FPGA
A42MX16
IN
A
105
LT
C
1410C
S
W
P
G
A
206U
D
G
407D
J
D
G
407D
J
D
G
407D
J
DC/DC CONVERTER
IN 5V OUT ±15V / ±65mA
DC/DC CONVERTER
IN 5V OUT ±5V / ±250mA
035
BIP
UNI
AIN32/12RB
ANALOG
INPUT
DIL switches
1
2
Voltage range
DIL1
DIL2
±10V bipolar (default)
BIP
BIP
0…10V unipolar
UNI
UNI
not allowed
BIP
UNI
not allowed
UNI
BIP
Fig. 79 –
: DIL switch settings for the input voltage range