Calibration with ADbasic Programs
Calibration with Trimmers
ADwin
198
ADwin-Pro
Hardware, manual version 2.9, June 2006
AIn 8/12
– AIn 8/12
1. Calibrate Offset:
• Enter the digital
min. test value
for
PAR_8
in the parameter
window and confirm with "
Send
" or by pressing
[RETURN]
.
• Set the voltage value with the offset trimmer for the relevant
channel.
2. Calibrate Gain:
• Enter the digital
max. test value
for
PAR_8
in the parameter
window and confirm with "
Send
" or by pressing
[RETURN]
.
• Set the voltage value with the corresponding gain trimming
potentiometer.
3. Check:
• Check all 3 test values from
AIn 32/12
– AIn 32/12 (bipolar und unipolar)
1. Calibrate Offset:
• Enter the digital
min. test value
for
PAR_8
in the parameter
window and confirm with "
Send
" or by pressing
[RETURN]
.
• Set the voltage value with the offset trimmer for the relevant
channel (BPO, UPO).
2. Calibrate Gain:
• Enter the digital
max. test value
for
PAR_8
in the parameter
window and confirm with "
Send
" or by pressing
[RETURN]
.
• Set the voltage value with the corresponding gain trimming
potentiometer.
3. Check:
• Check all 3 test values from
AIn F 8/12, AIn F 4/12
– AIn F 8/12, AIn F 4/12
1. Calibrate Offset:
• Enter the digital
mean test value
for
PAR_8
in the parameter
window and confirm with "
Send
" or by pressing
[RETURN]
.
• Set the voltage value with the offset trimmer for the relevant
channel.
2. Calibrate Gain:
• Enter the digital
max. test value
for
PAR_8
in the parameter
window and confirm with "
Send
" or by pressing
[RETURN]
.
• Set the voltage value with the corresponding gain trimming
potentiometer.
3. Check:
• Check all 3 test values from
AIn F 8/16, AIn F 4/16
– AIn F 8/16, AIn F 4/16
1. Calibrate Offset:
• Enter the digital
mean test value
for
PAR_8
in the parameter
window and confirm with "
Send
" or by pressing
[RETURN]
.
• Set the voltage value with the offset trimmer for the relevant
channel.
2. Calibrate Gain:
• Enter the digital
max. test value
for
PAR_8
in the parameter
window and confirm with "
Send
" or by pressing
[RETURN]
.
• Set the voltage value with the corresponding gain trimming
potentiometer.
3. Check:
• Check all 3 test values from