The module Pro-PT100 has 4 or 8 inputs for connecting platinum temperature
sensors of the type Pt 100. The maximum possible measurement range is
-200°C…+266°C, depending on the temperature sensor (see the data sheets
of the manufacturers, e.g. Betatherm, Ephy-Mess, Heraeus, Jomo, Omega,
Sensycon, etc.).
The amplifier outputs are connected to a LEMO-socket via multiplexer. The
output must be connected to an additional analog input module. The multi-
plexer can be set by an
ADbasic
instruction.
Measurements are done with 2, 3 or 4 wire technique (input circuit see
). Measuring method, zero point and gain are set via jumpers and trimmers
on the circuit board (
).
The measuring methods and the wiring between senso and Pro-PT100 mod-
ule is described on
Fig. 282 – Pro-PT100-x Rev. A: Block diagram
A conversion table is available for the conversion of the temperature values
into the corresponding integer values. You will find the conversion table in the
ADbasic
online help, topic "hardware information".
Inputs
4 or 8
Amplifier version
2, 3 or 4 wires
Multiplexer settling time
15µs
Max. measurement range
-200°C…+266°C
Accuracy
±0.2°C
Output voltage range
±10V to LEMO socket (A-OUT)
I
1
= I
2
1mA
Connector
Omega subminiature connector, type:
SMP-K-F
optional: 37-pin DSub socket
Module width
Pro-PT100-4
5HP wide / 1 slot
Pro-PT100-8-D
5HP wide / 1 slot
Pro-PT100-8
10HP wide / 2 slots
Fig. 283 – Pro-PT100-x: Specification
InAmp
G = 50
+
-
4
1
Sense +
3
Source -
2
Sense -
Pt
100
U
out
Ch1
+Ub
2L & 3L
2L
I
1
I
2
OpAmp
G = 2
-
+
5V
R1
R2
C1
C2
3L
MUX
Ch2
Ch3
Ch4
Ch5
Ch6
Ch7
Ch8
A-OUT