ALARM MESSAGES (CAUSES AND ACTIONS TO BE TAKEN)
MESSAGE
CAUSES
ACTION TO TAKE
NO ALARMS
All works regularly
---
[000] SYSTEM RESTART
---
---
[001] INTERNAL PS FAIL
Internal supply voltage error
Contact the service
[002] CLOCK NOT SET
System Clock not set
Set the system clock from the converter menu 13
(see also MCP function ).
[003] SD CARD FAILURE
SD card not found or unreadable
check and/or replace SD card
[005] F-RAM ERROR
Error writing / reading Flash-RAM
Contact the service
[006] EXCITATION ERROR
The excitation of the sensor coils resulting
from cable is interrupted
Check the connecting cables to the sensor.
[007] SIGNAL ERROR
The measure is strongly effected by external
noise or the cable connecting the converter to
the sensor is broken.
Check the status of the cables connecting the
sensor, the grounding connections of the devices and
the possible presence of noise sources.
[008] PIPE EMPTY
The measuring pipe is empty or the detection
system has not been properly calibrated.
Check whether the pipe is empty or repeat the empty
pipe calibration procedure.
[009] FLOW>MAX+
The flow rate is higher than the maximum
positive threshold set.
Check the maximum positive flow rate threshold set
and the process conditions.
[010] FLOW>MAX-
The flow rate is higher than the maximum
negative threshold set.
Check the maximum negative flow rate threshold set
and the process conditions.
[011] FLOW<MIN+
The flow rate is lower than the minimum
positive threshold set.
Check the minimum positive flow rate threshold set
and the process conditions.
[012] FLOW<MIN-
The flow rate is lower than the minimum
negative threshold set.
Check the minimum negative flow rate threshold set
and the process conditions.
[013] FLOW>FULL SCALE+
The flow rate is higher than the full scale
positive value set on the instrument.
Check the full scale positive value set on the
instrument and the process conditions.
[014] FLOW>FULL SCALE-
The flow rate is higher than the full scale
negative value set on the instrument.
Check the full scale negative value set on the
instrument and the process conditions.
[015] PULSE1>RANGE
The pulse generation output 1 of the device is
saturated and cannot generate the sufficient
number of impulses.
Set a bigger unit of volume or, if the connected
counting device allows it, reduce the pulse duration
value.
[016] PULSE2>RANGE
The pulse generation output 2 of the device is
saturated and cannot generate the sufficient
number of impulses.
Set a bigger unit of volume or, if the connected
counting device allows it, reduce the pulse duration
value.
[017] CALIBR.ERROR
Calibration Error
Contact the service
[018] SYSTEM FREQ.ERR
System Freq. Error
Contact the service
[019] B.DATA NOT INIT
Uninitialized data system
Contact the service
[020] FL.SENSOR ERROR
Flow rate sensor error
Contact the service
[021] BATTERY LOW
(Rechargeable) battery depleted
Contact the service to Replace the battery
[022] BATTERY V>MAX
Battery voltage (rechargeable)> max. Allowed
Contact the service to Replace the battery
[023] BATTERY I>MAX
Battery charge current> max. allowed
Contact the service to Replace the battery
[024] MAIN PS V.ERR
Main supply voltage (+ 5V) out of tolerance.
Contact the service
[025] USB VOLTAGE ERR
Voltage of USB connection out of tolerance.
Contact the service
[026] SDC ALMOST FULL
SD card space <500 MB.
For more information see function "12.9" page36.
[027] SDC FULL
SD card out of memory
Memory Full. You can not save logger. Contact the
service to replace the SD memory.
[028] BATT.TEMP.CRIT
The battery can not be charged. The
temperature is out of range ( detected
temperature <0 C° or temperature >50°)
Wait for the normal temperature reset. View
Environmental Use Conditions "Environmental Use