Delay before generating error. This function is useful to prevent unexpected lock to zero of measure caused by sporadic
events (empty pipe, excitation error, signal error)
Enable the Automatic sensor verification (see BIV optional function).
This feature appears only when the process conditions are as follow:
measure filters recommended (Damping) set to SMART 2 second / 5 second
stable flow rate and lower than 0.1% of the absolute scale (10 m/s)
It must have elapsed at least 10 minutes after the last significant change of flow rate
When the above conditions are met, the zero point calibration function will appears on the display, press the “Enter /
ESC” and automatically the procedure will start.
NOTICE: Be sure that the the sensor is completely full of liquid and perfectly still. Even subtle movements of the fluid
can cause significant errors, therefore proceed with great care.
Linearization coefficient for negative flow, reserved to the service. This command is only showed if SMODL = 000.
(POS. 1.24)
Signal Error ALarm Time
[S.err.delay=m xxx]
AL4
[SEALT]
(POS. 1.26) SET KL values
[Zero point cal.]
AL4
[SETKL]
(POS. 1.27 SET KJ values
[KL=XX +/- XXXXXXXXX]
AL4
[SETKJ]
(POS. 1.25)
Automatic Sensor VeriFy Enable
[Sens. verify= OFF]
AL3
[ASVFE]