The manuf
actur
er guar
antees only Engl
ish text a
vailable on our web si
te www
.isoil.com
MAN_MV110_EN_IT_IS_R13_1.04.XXXX
80 / 89
READING AND MEANING OF FILE STESTLOG.CSV
Registration Number
Data
Hour
Error code in hexadecimal format (0 = no error)
The temperature unit (degrees F or C) (#) (*)
CPU temperature
The unit of voltage CPU (V) (#)
Voltage measured on the electrode E1
The unit of voltage E1 (V) (#)
Reference voltage electrode E2
The unit of voltage E2 (V) (#)
Differential voltage E1-E2
The unit of voltage (V) (#)
Common mode voltage (E1 + E2) / 2
The unit of resistance (ohm) (#)
Resistance measured between E1 and the common
The unit of resistance (ohm) (#)
Resistance measured between E2 and the common
The unit of voltage (V) (#)
Common mode noise at low frequency
The unit of voltage (V) (#)
Differential mode noise at low frequency
The unit of voltage (mV) (#)
Mode ADC noise at low frequency differential
The unit of voltage (mV) (#)
Mode ADC noise high frequency differential
The unit of voltage (V) (#)
Analog circuitry positive supply voltage
The unit of voltage (V) (#)
Negative supply voltage analog circuits
The unit of current (mA) (#)
Excitation current of the coils
The unit of resistance (ohm) (#)
Measurement of the sensor coil resistance
The temperature unit (degrees F or C) (#) (*)
Temperature of the sensor coils
The unit of current (mA) (#)
The coil leakage current (insulation fault)
Unit of measure of time (ms) (#)
Rise time current phase A
Unit of measure of time (ms) (#)
Rise time current phase B
NOTE:
(#): The units are registered only if the appropriate function of the DATA LOGGER is active. Otherwise the field is empty.
(*): The temperature values can be expressed in degrees F or C, depending on the drive configuration