The manuf
actur
er guar
antees only Engl
ish text a
vailable on our web si
te www
.isoil.com
MAN_MV110_EN_IT_IS_R13_1.04.XXXX
61 / 89
Example: Sensor Verify
Example: Generic File
The function “Sensor Verify” creates the “STESTLOG.CSV” file, according to the following conditions:
1) Activate the SDC / RTC option in group “HW Config” (by factory see the order code)
2) Activate the “BIV” in the group “PRODUCT CODE” (by factory see the order code)
3) Activate the “Sens.verify” function in the “Sensor” menu
Activated such conditions, the converter will perform every hour a measure of sensor’s parameters and record a line of the file
“STESTLOG.CSV”; it is possible even a manual verification by the command “sens.verify” on the menu “Diagnostic” or through the
MCP command “SVERC”.
To download all the sensor measure for the current day in a specific
folder, set the below parameters as follows:
Data source:
Sensor Verify
When:
Today
Today:
All
Save path:
C: / .....
The parameters are set then click the Download button.
This function allows to select a file in the SD memory and
download it. In the “Generic file path” specify the file name
contained in the SD and indicate “Save path” where to save the
file. Set these values to proceed with the download by clicking
the “download”.