The MV 110 can show a 5 digits display for flow rate units; this mean the maximum flow rate value that can be
represented on the display is 99999 (no matter the positioning of the decimal point). The minimum is 0.0025.
The representable unit of measure depends on sensor flow rate and diameter; the permitted units are those, that
permits the instrument full scale value not exceeding 99999.
Example for DN 300, Full scale value: 3m/s:
PERMITTED unit of measure
(example): l/s (216.00); m3/h (777.60); m3/s (0.2160)
NOT PERMITTED unit of measure
(example): l/h (777600)
FLOW RATE VISUALIZATION
This symbol appears (red color on the virtual
display) only when the overall noise is over
2.5% of flow rate.
This FLAG becomes active when there is a flow variation (flow rate not stable).
The user has immediate access to the Quick Start menu when the converter is powered up by pressing the Enter key. If
access to the quick start menu does not occur, then it could be disabled using the function "9.11" page 35 .
QUICK START MENU
See programming
functions section
Access to all functions
Flow rate alert
The QUICK START MENU allows to user immediate access to some of the most commonly used functions; through MCP
software it possible customize this menu to make it suitable for the specific application.