83 / 89
CODE
ANOMALIES DESCRIPTION
ACTION TO TAKE
0000
NO ERROR
---
0001
SENSOR TEST INSULATION:
Generator power too low
Contact the service
0002
SENSOR TEST INSULATION:
Generator power too high
0004
SENSOR TEST INSULATION:
Phase 1 generator voltage too low
0008
SENSOR TEST INSULATION:
Phase 1 generator voltage too high
0010
SENSOR TEST INSULATION:
Phase 1 terminal voltage coils 1 too low
0020
SENSOR TEST INSULATION:
Phase 1 terminal voltage coils 2 too low
0040
SENSOR TEST INSULATION:
Phase 2 generator voltage too low
0080
SENSOR TEST INSULATION:
Phase 2 generator voltage too high
0100
SENSOR TEST INSULATION:
Phase 2 terminal voltage coils 1 too low
0200
SENSOR TEST INSULATION:
Phase 2 terminal voltage coils 2 too low
0400
SENSOR TEST INSULATION:
Insulation loss, leakage current out of tolerance
Check:
wiring between sensor converter
conditions of use
set parameters
If the problem persists contact the service
0800
TEST TEMPERATURE (RESISTANCE) COILS:
Temperature (resistance) out of tolerance
1000
TEST TIME GETTING ON CURRENT PHASE (A):
Value out of tolerance
2000
TEST TIME GETTING ON CURRENT PHASE (B):
Value out of tolerance
4000
TEST RESISTANCE INPUTS ELECTRODES:
Input value 1 out of tolerance
8000
TEST RESISTANCE INPUTS ELECTRODES:
Input value 2 out of tolerance
10000
SENSOR EXCITATION TEST:
Invalid sensor driving conditions
Contact the service
20000
REFERENCE VALUES TEST:
Invalid reference values
40000
ELECTRODES INPUT RESISTANCE TEST:
Empty pipe, test not possible
ERROR CODE TEST SYSTEM OF SENSOR
The codes are in hexadecimal format, the meaning is given for each bit. There are several possible error simultaneous
combinations (more bits active) then that will give the combined numerical codes.