The manuf
actur
er guar
antees only Engl
ish text a
vailable on our web si
te www
.isoil.com
MAN_MV110_EN_IT_IS_R13_1.04.XXXX
5 / 89
MODEL: Convert Model
S/N: Serial Number of the converter
SUPPLY: Main power supply
Hz: Supply frequency (AC)
POWER: Maximum power consumption
IP: Protection grade
T: Operation temperature
COUPLING: Serial number of sensor coupled
ITEM: Free for user
Power supply
versions
Power supply
voltage
Power supply
frequency
Min
Power
Max power
HV
100-240V~
45-66HZ
1,5 W
(Sensor only)
12 W
(all Loads)
LV
24-36V
//
24-36V~
45-66HZ
LLV
12-48V
//
The instrument label contain the following information:
The converter can be installed internally or externally
Altitude: untill 4000m (from -656 to 6560 feet)
Humidity range: 0-100%
If the converter is supplied in compact version (converter over the sensor), consider the ambient temperatures more
restrictive, otherwise refer to the relevant manuals.
* For discontinuous use, a thermostat heat source installation may be necessary.
TECHNICAL CHARACTERISTICS
Electrical Characteristic
Environmental Use Conditions
Data Plate
ATTENTION!
The battery will not be charged
outside the below limits :
T board MV110 < 0 °C
T board MV110 > 50 °C
AMBIENT TEMPERATURE
Aluminium
Reinf. Nylon
Min*
Max
Min*
Max
°C
-20
+ 60
-10
+ 50
°F
-4
+140
+14
+122
Converter classification: class I,
IP67 (where: 6 = totally protected against dust, 7 = Protected against the effects of temporary immersion) /
IP68 (where: 6 = totally protected against dust, 8 = continuous immersion 1.5 m; 1 h)
for aluminum and PA6 housing, installation category (overvoltage) II, rated pollution degree 2.
Voltage variations must not exceed ±10% of the nominal one.
Input/output are insulated
The output 4-20mA (optional) is electrically connected to the ON/OFF outputs and the output power supply (24V
).
Version LV/LLV : inrush current < 20A
Version HV : inrush current < 25A
Note:
The devices powered at 24-36 V and 12-48 V must only be powered with power supplies compliant with the
IEC61010 standard